CVE-2018-14781

Severity CVSS v4.0:
Pending analysis
Type:
Unavailable / Other
Publication date:
13/08/2018
Last modified:
22/05/2025

Description

Medtronic MiniMed MMT <br /> <br /> devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Vulnerable products and versions

CPE From Up to
cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:*
cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:*