CVE-2024-48971
Severity CVSS v4.0:
Pending analysis
Type:
CWE-798
Use of Hard-coded Credentials
Publication date:
14/11/2024
Last modified:
15/11/2024
Description
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Impact
Base Score 3.x
9.30
Severity 3.x
CRITICAL