CVE-2015-3952
Severity CVSS v4.0:
Pending analysis
Type:
CWE-200
Information Leak / Disclosure
Publication date:
25/03/2019
Last modified:
09/10/2019
Description
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
5.00
Severity 2.0
MEDIUM
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:* | 13.4 (including) | |
cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:* | 13.6 (including) | |
cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:* | 3.13 (including) | |
cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page