CVE-2017-12716
Severity CVSS v4.0:
Pending analysis
Type:
CWE-319
Cleartext Transmission of Sensitive Information
Publication date:
25/04/2018
Last modified:
09/10/2019
Description
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Impact
Base Score 3.x
6.50
Severity 3.x
MEDIUM
Base Score 2.0
3.30
Severity 2.0
LOW
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:* | f0b.0e.7e (excluding) | |
| cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:* | ||
| cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:* | f0b.0e.7e (excluding) | |
| cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:* | ||
| cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:* | f10.08.6c (excluding) | |
| cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:* | ||
| cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:* | f10.08.6c (excluding) | |
| cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page