CVE-2017-12720
Severity CVSS v4.0:
Pending analysis
Type:
CWE-306
Missing Authentication for Critical Function
Publication date:
15/02/2018
Last modified:
03/10/2019
Description
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.
Impact
Base Score 3.x
8.10
Severity 3.x
HIGH
Base Score 2.0
6.80
Severity 2.0
MEDIUM
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:* | ||
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:* | ||
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:* | ||
| cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page