CVE-2020-25183
Severity CVSS v4.0:
Pending analysis
Type:
CWE-287
Authentication Issues
Publication date:
14/12/2020
Last modified:
22/05/2025
Description
Medtronic MyCareLink Smart 25000 contains <br />
<br />
an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Impact
Base Score 3.x
8.00
Severity 3.x
HIGH
Base Score 2.0
5.80
Severity 2.0
MEDIUM
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:* | ||
cpe:2.3:h:medtronic:mycarelink_smart_model_25000:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page