CVE-2020-27262
Severity CVSS v4.0:
Pending analysis
Type:
CWE-79
Cross-Site Scripting (XSS)
Publication date:
08/01/2021
Last modified:
14/01/2021
Description
Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 A stored cross-site scripting (XSS) vulnerability exists in the affected products that allow an attacker to inject arbitrary web script or HTML via the filename parameter to multiple update endpoints of the administrative web interface.
Impact
Base Score 3.x
5.40
Severity 3.x
MEDIUM
Base Score 2.0
3.50
Severity 2.0
LOW
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:innokasmedical:vital_signs_monitor_vc150_firmware:*:*:*:*:*:*:*:* | 1.7.15 (excluding) | |
| cpe:2.3:h:innokasmedical:vital_signs_monitor_vc150:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page