CVE-2021-33886
Severity CVSS v4.0:
Pending analysis
Type:
CWE-134
Format String Vulnerability
Publication date:
25/08/2021
Last modified:
12/07/2022
Description
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Impact
Base Score 3.x
8.80
Severity 3.x
HIGH
Base Score 2.0
5.80
Severity 2.0
MEDIUM
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:* | 012u000062 (excluding) | |
| cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:* | ||
| cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page