CVE-2021-37166

Severity CVSS v4.0:
Pending analysis
Type:
CWE-120 Buffer Copy without Checking Size of Input ('Classic Buffer Overflow')
Publication date:
02/08/2021
Last modified:
07/11/2023

Description

A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.

Vulnerable products and versions

CPE From Up to
cpe:2.3:o:swisslog-healthcare:hmi-3_control_panel_firmware:*:*:*:*:*:*:*:* 7.2.5.7 (excluding)
cpe:2.3:h:swisslog-healthcare:hmi-3_control_panel:-:*:*:*:*:*:*:*