CVE-2021-43355
Severity CVSS v4.0:
Pending analysis
Type:
CWE-287
Authentication Issues
Publication date:
21/01/2022
Last modified:
09/08/2022
Description
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges.
Impact
Base Score 3.x
9.80
Severity 3.x
CRITICAL
Base Score 2.0
7.50
Severity 2.0
HIGH
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:* | 3.3.0 (including) | |
cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:* | ||
cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:* | ||
cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:* | ||
cpe:2.3:o:fresenius-kabi:agilia_connect_firmware:*:*:*:*:*:*:*:* | d25 (including) | |
cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:*:*:*:*:*:*:*:* | 3.0 (excluding) | |
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:-:*:*:*:*:*:* | ||
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:* | ||
cpe:2.3:h:fresenius-kabi:link\+_agilia:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page