CVE-2014-5403

Severity CVSS v4.0:
Pending analysis
Type:
CWE-321 Use of Hard-coded Cryptographic Key
Publication date:
03/04/2015
Last modified:
03/11/2025

Description

Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network.

Vulnerable products and versions

CPE From Up to
cpe:2.3:a:hospira:mednet:*:*:*:*:*:*:*:* 5.8 (including)