CVE-2015-1012
Severity CVSS v4.0:
Pending analysis
Type:
CWE-200
Information Leak / Disclosure
Publication date:
25/03/2019
Last modified:
09/10/2019
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
5.00
Severity 2.0
MEDIUM
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:pfizer:lifecare_pca_infusion_system_firmware:*:*:*:*:*:*:*:* | 5.0 (including) | |
| cpe:2.3:h:pfizer:lifecare_pca_infusion_system:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page