CVE-2015-3956
Severity CVSS v4.0:
Pending analysis
Type:
CWE-345
Insufficient Verification of Data Authenticity
Publication date:
25/03/2019
Last modified:
09/10/2019
Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Impact
Base Score 3.x
9.80
Severity 3.x
CRITICAL
Base Score 2.0
10.00
Severity 2.0
HIGH
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:* | 13.4 (including) | |
cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:* | 13.6 (including) | |
cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:* | 3.13 (including) | |
cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page