CVE-2016-5085
Severity CVSS v4.0:
Pending analysis
Type:
CWE-330
Use of Insufficiently Random Value
Publication date:
05/10/2016
Last modified:
12/04/2025
Description
Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then engaging in an authentication handshake.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
7.80
Severity 2.0
HIGH
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:animas:onetouch_ping_firmware:-:*:*:*:*:*:*:* | ||
| cpe:2.3:h:animas:onetouch_ping:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page
References to Advisories, Solutions, and Tools
- http://www.kb.cert.org/vuls/id/884840
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRS
- http://www.securityfocus.com/bid/93351
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pump
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01
- http://www.kb.cert.org/vuls/id/884840
- http://www.kb.cert.org/vuls/id/BLUU-A9SQRS
- http://www.securityfocus.com/bid/93351
- https://community.rapid7.com/community/infosec/blog/2016/10/04/r7-2016-07-multiple-vulnerabilities-in-animas-onetouch-ping-insulin-pump
- https://ics-cert.us-cert.gov/advisories/ICSMA-16-279-01