CVE-2018-10596
Severity CVSS v4.0:
Pending analysis
Type:
Unavailable / Other
Publication date:
03/07/2018
Last modified:
22/05/2025
Description
Medtronic 2090 CareLink Programmer <br />
<br />
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Impact
Base Score 3.x
7.10
Severity 3.x
HIGH
Base Score 2.0
5.20
Severity 2.0
MEDIUM
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:* | ||
cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page