CVE-2019-10966

Severity CVSS v4.0:
Pending analysis
Type:
CWE-287 Authentication Issues
Publication date:
10/07/2019
Last modified:
02/10/2020

Description

In GE Aestiva and Aespire versions 7100 and 7900, a vulnerability exists where serial devices are connected via an added unsecured terminal server to a TCP/IP network configuration, which could allow an attacker to remotely modify device configuration and silence alarms.

Vulnerable products and versions

CPE From Up to
cpe:2.3:o:ge:aestiva_7100_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:ge:aestiva_7100:-:*:*:*:*:*:*:*
cpe:2.3:o:ge:aestiva_7900_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:ge:aestiva_7900:-:*:*:*:*:*:*:*
cpe:2.3:o:ge:aespire_7100_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:ge:aespire_7100:-:*:*:*:*:*:*:*
cpe:2.3:o:ge:aespire_7900_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:ge:aespire_7900:-:*:*:*:*:*:*:*