CVE-2020-12036
Severity CVSS v4.0:
Pending analysis
Type:
CWE-319
Cleartext Transmission of Sensitive Information
Publication date:
29/06/2020
Last modified:
14/07/2020
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
5.00
Severity 2.0
MEDIUM
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:* | ||
cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:* | ||
cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:* | 3.0 (excluding) | |
cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page