CVE-2020-12048

Severity CVSS v4.0:
Pending analysis
Type:
CWE-319 Cleartext Transmission of Sensitive Information
Publication date:
29/06/2020
Last modified:
16/07/2020

Description

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Vulnerable products and versions

CPE From Up to
cpe:2.3:o:baxter:phoenix_x36_firmware:3.36:*:*:*:*:*:*:*
cpe:2.3:o:baxter:phoenix_x36_firmware:3.40:*:*:*:*:*:*:*
cpe:2.3:h:baxter:phoenix_x36:-:*:*:*:*:*:*:*


References to Advisories, Solutions, and Tools