Description
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
5.00
Severity 2.0
MEDIUM
Vulnerable products and versions
CPE | From | Up to |
---|---|---|
cpe:2.3:o:baxter:phoenix_x36_firmware:3.36:*:*:*:*:*:*:* | ||
cpe:2.3:o:baxter:phoenix_x36_firmware:3.40:*:*:*:*:*:*:* | ||
cpe:2.3:h:baxter:phoenix_x36:-:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page