CVE-2020-15486
Severity CVSS v4.0:
Pending analysis
Type:
Unavailable / Other
Publication date:
26/08/2020
Last modified:
21/07/2021
Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Impact
Base Score 3.x
6.50
Severity 3.x
MEDIUM
Base Score 2.0
3.30
Severity 2.0
LOW
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:o:drtrust:electrocardiogram_pen_firmware:2.00.08:*:*:*:*:*:*:* | ||
| cpe:2.3:h:drtrust:electrocardiogram_pen:2.00.08:*:*:*:*:*:*:* |
To consult the complete list of CPE names with products and versions, see this page