CVE-2021-32966
Severity CVSS v4.0:
Pending analysis
Type:
CWE-319
Cleartext Transmission of Sensitive Information
Publication date:
25/05/2022
Last modified:
08/06/2022
Description
Philips Interoperability Solution XDS versions 2.5 through 3.11 and 2018-1 through 2021-1 are vulnerable to clear text transmission of sensitive information when configured to use LDAP via TLS and where the domain controller returns LDAP referrals, which may allow an attacker to remotely read LDAP system credentials.
Impact
Base Score 3.x
7.50
Severity 3.x
HIGH
Base Score 2.0
4.30
Severity 2.0
MEDIUM
Vulnerable products and versions
| CPE | From | Up to |
|---|---|---|
| cpe:2.3:a:philips:interoperability_solution_xds:*:*:*:*:*:*:*:* | 2.5 (including) | 3.11 (including) |
| cpe:2.3:a:philips:interoperability_solution_xds:*:*:*:*:*:*:*:* | 2018-1 (including) | 2021-1 (including) |
To consult the complete list of CPE names with products and versions, see this page