Instituto Nacional de ciberseguridad. Sección Incibe
Instituto Nacional de Ciberseguridad. Sección INCIBE-CERT

CVE-2025-14096

Gravedad CVSS v3.1:
ALTA
Tipo:
No Disponible / Otro tipo
Fecha de publicación:
17/12/2025
Última modificación:
17/12/2025

Descripción

*** Pendiente de traducción *** A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.<br /> <br /> Other related CVE&amp;#39;s are CVE-2025-14095 &amp; CVE-2025-14097.<br /> <br /> <br /> <br /> Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.<br /> <br /> <br /> <br /> <br /> <br /> Required Configuration for Exposure:<br /> <br /> <br /> Attacker requires physical access to the analyzer.<br /> <br /> <br /> <br /> Temporary work Around:<br /> Only authorized people can physically access the analyzer.<br /> <br /> Permanent solution:<br /> Local Radiometer representatives will contact all affected customers to discuss a permanent solution.<br /> <br /> Exploit Status:<br /> <br /> <br /> Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.