Instituto Nacional de ciberseguridad. Sección Incibe
Instituto Nacional de Ciberseguridad. Sección INCIBE-CERT

CVE-2019-25721

Gravedad CVSS v4.0:
ALTA
Tipo:
CWE-400 Consumo de recursos no controlado (Agotamiento de recursos)
Fecha de publicación:
02/06/2026
Última modificación:
17/06/2026

Descripción

*** Pendiente de traducción *** Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.