CVE-2021-4479
Gravedad CVSS v4.0:
MEDIA
Tipo:
No Disponible / Otro tipo
Fecha de publicación:
02/06/2026
Última modificación:
10/06/2026
Descripción
*** Pendiente de traducción *** Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
Impacto
Puntuación base 4.0
6.30
Gravedad 4.0
MEDIA
Puntuación base 3.x
4.00
Gravedad 3.x