Instituto Nacional de ciberseguridad. Sección Incibe
Instituto Nacional de Ciberseguridad. Sección INCIBE-CERT

CVE-2025-53963

Gravedad CVSS v3.1:
CRÍTICA
Tipo:
No Disponible / Otro tipo
Fecha de publicación:
04/12/2025
Última modificación:
05/12/2025

Descripción

*** Pendiente de traducción *** An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.